Theory


 

The amount of light scattered is directly proportional to the product of the weight-average molar mass and the macromolecule (solute) concentration, i.e., LS ~ Mw·c (Wyatt, 1993).

Based on Zimm’s formalism, the Rayleigh-Debye-Gans light scattering model for dilute polymer solutions can be expressed as equation [1] (Wyatt, 1993):
 

                     [1]
 

where:
 

The function P(Q) describes the angular dependence of scattered light.

The expansion of 1/ P(Q) to first order gives:
 

1/ P(Q) = 1 + (16p2/3l2) <rg2>. sin2(Q/2) + f4 sin4(Q/2) +...
 

At low angles the angular dependence of light scattering depends only on the mean square radius <rg2> (alternatively called radius of gyration) and is independent of molecular conformation or branching.   A  plot  of   K*c /R(Q) vs. sin2(Q/2) (Zimm plot) yields a curve whose intercept gives Mw and whose slope at low angles gives <rg2>.



MW calculation based on the"Two detector" approximation

During gel permeation chromatography, the protein concentrations are so low that the viral coefficient term is negligible (Wen et al, 1996).  In addition, the term (16 p2/3l2) <rg2> sin2(Q/2) will be negligible if we measure the light scattering at a sufficiently small angle.  Furthermore, this term will be negligible at all angles for <rg2> smaller then 15 nm;  thus, the scattered light signal  measured at 90o is commonly used in this approximation (Takagi, 1990;  Wen et al, 1996).  This approximation is true for folded polypeptides with molecular weight (MW) smaller than about 5x107 Da.  Under such conditions, Eq. [1] can be simplified to:

 (K* c)/R(Q ) = 1/M                             [2]

Substituting K* with 4p2n2 (dn/dc)2 / (lo4NA) and introducing KLS (calibration constant for LS detector) the intensity of scattered light is given by:

 (LS) = KLS c M (dn/dc)2                     [3]

Similarly, the refractive index signal, (RI), can be expressed as:

(RI) = KRI c (dn/dc)                             [4]

where KRI is the instrument calibration constant.
 

For proteins/protein complexes that contain only polypeptides (no carbohydrates) the dn/dc is constant (~0.19 ml/g) and nearly independent of amino acid composition.  Thus, we can determine the concentration of protein from the RI signal and the MW from the ratio of the two detectors, (LS) and (RI):

 MW = K’ (LS)/(RI)                             [5]

 where K’ = KRI/[KLS(dn/dc)].

This is the so-called "two-detector" method.  This method is widely used but it is valid only when dn/dc is known (Takagi, 1990, Wen et al, 1996).  In this approach the MW is NOT determined from the absolute light scattering measurements but is calculated from Eq. [5] assuming K’ is known.  The instrument calibration constant, K’, is derived from running protein standards.
 



Absolute MW Calculation based on the ASTRA software

Alternatively, the MW can be determined directly from the absolute light scattering measurements by solving Eq. [1] using the ASTRA software (note that the concentration needs to be measured independently).  Click here fordetails on ASTRA calculations.
 

References:

1. Wyatt, P. J. (1993) Anal. Chim. Acta, 272, 1-40

2. Takagi, T. (1990) J. Chromatogr. , 506, 409-416

3. Wen, J., Arakawa, T. & Philo, J. S. (1996) Anal. Biochem., 240, 155-166