The new service offers the exciting capability of determining the native
molecular weights of proteins, their protein:protein and other macromolecular
complexes. The
MW determination depends only upon the downstream light scattering (LS)
and refractive index (RI) detectors, and is thus completely independent
of elution position.Hence,
non-globular shape and/or possible interaction with the size exclusion
chromatography (SEC) support, which otherwise pose severe limitations with
regard to the use of SEC for estimating MW, have no impact on the molecular
weights determined by LS. Based on analyzing
15 protein standards that range from 6.5 kDa to 457 kDa, native MW’s
may be determined quickly and with an error of less than 10%.
The size exclusion chromatography is carried out on one of the following columns:
Amount: 100-300 µg
protein/run with the larger amounts being required as the MW is decreased
below about 40 kDa
(click
here for detailed table for sample requirements for proteins)
PLEASE NOTE: The
amount indicated refers to FILTRATED sample injected onto the column.
CAUTION:
Please note that all samples are filtered through a 0.22 µm filter
(low binding Durapore® membrane, Millipore, Cat. No. UFC30GVNB) immediately
prior to analysis. We strongly recommend that a trial filtration
is performed before sample submission to avoid/account for sample loss
during this procedure.
Note: The filtration is performed just before injecting the sample on the system. At this point the system is already equilibrated and calibrated at the conditions requested by the investigator. Hence, the set-up charge will be billed even for samples that failed to filter through the 0.22 µm filter (low binding Durapore® membrane, Millipore, Cat. No. UFC30GVNB) and subsequently were not injected/analyzed.
Volume: 250 µl or less is recommended; up to 750 µl may be loaded with some loss of resolution
Glycerol: less than 5%*
Please contact Ewa Folta-Stogniew at least 48 hours in advance before submitting samples (especially unstable proteins) to ensure system availability.
* samples can be submitted in buffer with higher glycerol concentrations and buffer exchange carried out prior to the analysis for an additional charge
For sample submission form go to HPLC
SEC/Laser Light Scattering Sample Submission Page.
For pricing go toService
Charges for HPLC SEC/Laser Light Scattering.
|
|
|
|
|
|
Front Page |
|
|
|
|
|
|
|