ASTRA Summary Report Explanation Chart


 

The ASTRA Summary Report contains information about the calculated molecular weights (MW) as well as collection and processing parameters that were applied during data analysis.

The following chart is designed to guide you through the Summary Report (adopted from ASTRA manual; Wyatt Technology).  Please note that the Result section presents data for ALL peaks selected for analysis.  Peak selection and boundaries are shown on the attached "Peak ID" graph.
 

COLLECTION INFORMATION

This part of the report contains information about data collection, instrument used, type of  flow cell in the LS detector and solvent in the mobile phase.

The values of the calibration constants used during data processing are shown along with the information regarding mass detector (refractive index or UV) was used to estimate eluted mass:
 

AUX1: UV detector (which might be used as a mass detector when the extinction coefficient is known)

AUX2:  RI detector (might be used as a mass detector when the dn/dc value is known)

Flow rate used during SEC run is also shown.

 

PROCESSING INFORMATION

 
This part of the report refers to the fitting method that was used by ASTRA to carry out the molecular mass determination

Processing time:
        date and time of processing
 

Fit method/model:

Calculation method: dn/dc value + AUX constant: means that the known dn/dc was used along with the AUX detector reading. Alternatively, the total injected mass can be used with the assumption of 100% recovery of that mass within the analyzed peak value (dn/dc is a refractive index increment of the solute, which usually means the increase in refractive index with protein concentration). Detectors used: the subset of signals from LS detectors used in the analysis (there is a total of 18 detectors in the DAWN instrument but for aqueous solutions the signal from lower angle detectors, i.e. detectors #3-6 and #16-18, is very noisy and for some samples these readings may be deleted during data processing)   RESULTS

This section summarizes the molecular weight calculations for each peak selected (please refer to attached "Peak ID" graph for peak identification).

Volume:

Slices: A2: Fit Degree: Injected Mass: Calc. Mass: dn/dc value:  Polydispersity (Mw/Mn): indicates whether the peak is homogenous with respect to molar mass.  A homogeneous (i.e., monodisperse) sample is one that contains only one type of molecule of defined MW (e.g. pure monomeric protein), thus the average mass is independent of the averaging method and polydispersity will equal 1. If the sample contains a mixture of species of different molar masses (i.e., polydisperse sample; for example, a mixture of dimer or tetramer and monomer) the average mass will depend on the averaging method and the polydispersity will be different from 1. Polydispersity (Mz/Mn): Molar Mass Moments (g/mol)
   
 

Mn is the molar mass (or molecular weight, MW); this might for instance be measured by osmometry.

 

 

Mw: Weight-average molar mass is defined as:

 

 
Mw is the molar mass (or molecular weight, MW); this might for instance be measured by light scattering method.

   

 
 

Mz: Z-average molar mass is defined as:
 
  Mz is the molecular weight (MW); this might for instance be measured during sedimentation analysis.
   
R.M.S. Radius Moments (nm)
In general, ASTRA is solving the following equation that relates the excess scattered light to the concentration and weight-average molar mass:

 

                               (1)

 

where:

 
R(Q)     is the excess intensity of scattered light at DAWN angle Q

c           is the sample concentration

Mw       is the weight-average molecular weight

A2           is a second virial coefficient

K*        is an optical parameter equal to 4p2n2 (dn/dc)2 / (lo4NA)

n           is the solvent refractive index and dn/dc is the refractive index increment

NA         is Avogadro’s number

lo          is the wavelength of the scattered light in vacuum.

 

The function P(Q) describes the angular dependence of scattered light.

The expansion of 1/ P(Q) to first order gives:

 

1/ P(Q) = 1 + (16p2/3l2) <rg2>. sin2(Q/2) + f4 sin4(Q/2) +...

 

There are several ways in which Eq. (1) (that relates excess of scattered light to concentration and weight-average molar mass) can be solved to get Mw and <rg2>. The parameter "Fit method/model" refers to the actual method applied.

 

Debye Fit Method:

constructs a plot of R(Q)/K*c vs. sin2(Q/2) and fits a polynomial in sin2(Q/2) to the data, thereby obtaining Mw and <rg2> from the intercept and slope at zero angle (the order of polynomial used during fitting is given in the RESULT section as "FIT DEGREE").

It gives good results over a wider range of molecular weight as compared to the Zimm formalism.

 

Zimm Fit Method: constructs a plot of K*c/R(Q) vs. sin2(Q/2) and fits a polynomial in sin2(Q/2) to the data, thereby obtaining Mw and <rg2> from the intercept and slope at zero angle (the order of polynomial used during fitting is given in the RESULT section as "FIT DEGREE").

 
This is the most popular method for analysis of light scattering data. It works well for mid-sized molecules (rms radius up to ~20-50 nm).

 

Berry Fit Method: constructs a plot of SQRT{K*c/R(Q)} vs. sin2(Q/2) and fits a polynomial in sin2(Q/2) to the data  (the order of polynomial used during fitting is given in RESULT section as "FIT DEGREE").

 
Useful for large molecules.

 

Random Coil Method:
 
Eq. (1) is solved after inserting the theoretical form factor P(Q) for random coils given by:

where u = (4p/l)2<r2>sin2(Q/2).

P(Q) is a nonlinear function of <r2>, an iterative nonlinear least square fit is used during fitting. This fitting method might be advantageous for large random coil molecules  (the order of polynomial used during fitting is given in RESULT section as "FIT DEGREE").

 

Please note that all the above methods will give nearly identical results for small molecules (rms radius <10nm). Thus, any of the fitting formalisms can be used interchangeably for proteins with MW <1x106 Da.