LAB
NEWS
September
1996 . . . . . . . . . . Vol. 38 No. 1
Chairman:
Peter Jatlow,
M.D.
Editors: Edward L. Snyder, M.D.; Petrie M. Rainey, M.D.,
Ph.D.
Production Assistant: Terri M. Fiondella
Contributors: Peter Marone, MT (ASCP), MBA; Petrie Rainey,
M.D., Ph.D.; Henry M. Rinder, M.D.; Brian Smith, M.D.
POINT
OF CARE TESTING: A PRIMER
Point
of care testing (POCT) is performance of a laboratory test at or near
the patient's bedside. POCT has recently become an area of high interest,
as the result of technological innovation and government regulation.
Thanks to computer chips, there is a growing stream of sophisticated,
miniaturized instruments that allow various kinds of testing at the
point of care that previously were only available in central laboratories.
At the same time, regulatory agencies, such as the Joint Commission
for the Accreditation of Healthcare Organizations (JCAHO), have begun
enforcing a series of Federal regulations that apply to all forms
of medical laboratory testing, including POCT. Yale-New Haven Hospital
recently established a Point of Care Testing Committee to assure the
quality of patient testing done at the point of care. The Committee
will also advise the Hospital on the introduction of new POCT devices.
In addition, the position of Point of Care Coordinator was created.
The Coordinator's most important responsibility is to advise and assist
those who perform POCT in complying with government regulations and
JCAHO requirements.
Government
Regulation
The Clinical Laboratory Improvement Amendments of 1988, often referred
to as CLIA '88, or simply CLIA, were passed to regulate all clinical
laboratory testing (previously only testing in central laboratories
had been regulated). The regulations implementing the law were published
in 1992 and took full effect in 1994. There was considerable publicity
about the regulations when they came out, particularly with regard
to their impact on physician's office laboratories. The emphasis
on office laboratories may have led some to fail to realize that
the regulations applied to all testing at any site (including, for
example, stool for occult blood or urine dipsticks).
According to these regulations, laboratory testing is any examination
or testing of any specimen obtained from a patient for the purposes
of medical diagnosis, prognosis or monitoring, whether or not it
takes place in an actual laboratory. (When no specimen is involved,
such as occurs when measuring the blood pressure or using a pulse
oximeter, this is considered patient monitoring and is not regulated.)
The only testing exempted from the regulations was forensic testing,
drug abuse testing and pure research testing. (The Yale Human Investigation
Committee has determined that the exemption does not apply
to research testing where the results are reported into the medical
record or used to make medical management decisions. Such tests
must adhere to all applicable regulations.) The State of Connecticut
has also passed state regulations that closely follow the Federal
ones.
The extent of regulation depends on the complexity of the tests.
A limited number of simple tests are referred to as "waived"
tests, because they are free of extensive regulation (see Table
1). Tests done with commercial kits or devices that require
limited operator training are usually classified as "moderately
complex." To qualify as moderately complex, testing must adhere
to all manufacturer's instructions. The regulations for moderately
complex tests are substantial. All other tests are considered "highly
complex" and are the most intensively regulated.
There is a special category of moderately complex testing known
as provider-performed microscopy (PPM). These are microscopic examinations
of certain patient specimens performed as part of a physical examination
(see Table 2). When performed by a "laboratory"
that does not perform any other moderately or highly complex testing,
PPM is treated similarly to waived testing. In particular, it is
exempt from the requirement for a biannual inspection.
One universal requirement, common to all categories of testing,
is that the testing be carried out under the authority of an appropriate
certificate issued by the Health Care Finance Administration. All
persons who perform point of care testing should assure that their
testing is covered by an appropriate certificate. (In general, nurses
and other hospital employees will be covered by certificates held
by the Department of Nursing or the Department of Laboratory Medicine;
physicians who perform testing personally should be covered by certificates
held individually or by their practice groups.)
The JCAHO requirements for waived testing are slightly more demanding
than the Federal requirements and are shown in Table
3. Table 4 summarizes the major requirements
for moderately complex testing. Highly complex testing is generally
not appropriate for performance at the point of care.
New
POCT Devices
The bedside blood glucose monitor is the prototype for many of the
new POCT instruments. In the past few years, devices have been introduced
for measuring blood gases, electrolytes, BUN, hematocrit, activated
clotting times and a variety of other analytes. Instruments are
available from several manufacturers for most of these tests. Another
category of POCT devices is self-contained immunoassays giving a
simple positive or negative result. Urine pregnancy tests are the
prototypes for these assays. More recent entrants include devices
for drug abuse testing. Many other instruments and devices are in
the pipeline. Such devices are commonly marketed in a manner similar
to "me-too" drugs with little in the way of meaningful
performance evaluations or comparisons. Even less information is
available on whether use of these devices leads to actual improvement
in patient care, when compared with testing in a central laboratory.
The assumption is that faster is always better. While this will
certainly be true in some cases, it may not be in others. The only
careful, prospective studies of the impact of POCT devices have
come from Washington University. In one study, the use of POC coagulation
testing during cardiopulmonary bypass surgery resulted in substantial
reductions in O.R. time and in blood product use. In another, however,
introduction of a POCT device for measuring electrolytes, BUN and
glucose in the Emergency Department actually resulted in a slight
increase in the average length of stay.
The clever technology that makes these instruments possible must
be paid for. The result is that most test cartridges are relatively
expensive, often more than 10 times the cost of reagents used for
centralized testing. Medical staff time is also required to perform
the test and complete the required documentation. For moderately
complex tests, several minutes per result may be needed. The reduction
in the volume of testing in the central lab due to POCT is typically
small and may be nonexistent. The volume of blood glucose testing
in the YNHH Clinical Chemistry Laboratory actually grew throughout
the entire period during which bedside blood glucose monitoring
was introduced. Because there is no meaningful reduction in the
need for laboratory staff or instrumentation, savings in the central
lab due to avoided tests is limited to the cost of the reagents.
The sometimes considerable increase in expenses associated with
POCT should be justified by clearly demonstrable improvements in
patient care.
Point
of Care Testing Committee and Coordinator
Because of the increasing importance of POCT, Yale-New Haven Hospital
recently established a Point of Care Testing Committee. The primary
mission of the POCT Committee is to assure the quality of POCT carried
out at YNHH. While this includes assuring that POCT is done in compliance
with regulations, the more important responsibility is assuring
the achievement of the objective of these regulations, the production
of accurate and reliable results. A related responsibility is assuring
that POCT is carried out in a consistent manner throughout the hospital
and that there is a single standard of quality for all laboratory
testing, regardless of location.
A further responsibility is to advise the Hospital on the introduction
of new POCT devices. In this capacity, the POCT Committee will function
somewhat similarly to the Pharmacy and Therapeutics Committee, which
reviews requests to add new drugs to the formulary. A major difference
is that approval of a device by the POCT Committee does not guarantee
its introduction. Approved POCT devices must also be successful
in a competitive review with other proposed new programs. Persons
wishing to propose a device for consideration should obtain an application
form from the POCT Coordinator.
At the same time the POCT Committee was formed, the position of
POCT Coordinator was also created. Medical technologist Peter Marone,
MT(ASCP), MBA, currently holds the position. He can be reached by
campus mail (CB 407), phone (5-5212), pager (128*2351) and e-mail
(peter.marone@yale.edu). Responsibility for correct test performance
and regulatory compliance lies with the personnel performing the
tests and their supervisors. The POCT Coordinator can provide advice
and limited assistance in meeting these responsibilities (There
are several thousand medical staff who perform POCT and one coordinator.)
The first major task being undertaken by the Coordinator is to define
the extent of POCT in the Hospital and its clinics. It is possible
that not all POCT has yet been identified. If you are performing
any form of patient testing and there have been no changes made
recently to comply with the regulations, you should notify Mr. Marone.
Any other questions regarding POCT raised by this article should
also be addressed to Mr. Marone.
This brief overview is of necessity incomplete. The full regulations
encompass several hundred pages in the Federal Register. Moreover,
the regulations are continuing to evolve. For specific and up-to-date
information, please contact the Laboratory Medicine Regulatory Compliance
Specialist, John Chase, PhD (5-7154), or the POCT Coordinator, Peter
Marone (5-5212).
References
- Department
of Health and Human Services. Clinical Laboratory Improvement
Amendments of 1988; Final Rule (42 CFR Part 405, et al.). Federal
Register, 1992; 57:7001-7288.
- Despotis
GJ, et al. The impact of heparin concentration and activated clotting
time monitoring on blood conservation. A prospective, randomized
evaluation in patients undergoing cardiac operation. J Thorac
Cardiovasc Surg 1995;110:46-54.
- Parvin
CA, et al. Impact of point-of-care testing on patients' length
of stay in a large emergency department. Clin Chem 1996;42:711-717.
Petrie M. Rainey, M.D., Ph.D. Peter Marone, MT(ASCP), MBA
Table 1
| Table 1. WAIVED TESTS
|
|---|
| (These are simple tests that are minimally regulated -- See
Table 3)
- Urinalysis dipsticks*
- Whole blood glucose*
- Hemoccult*
- Gastroccult*
- Urine pregnancy test (visual color comparison only)*
- pH testing for body fluids other than blood (qualitative color
comparisons only)*
- Ovulation test (visual color comparison only)
- Erythrocyte sedimentation rate (non-automated)
- Hemoglobin-copper sulfate (non-automated)
- Hemoglobin by single analyte instruments
- Spun microhematocrit
- Rapid screen for Group A Streptococcus. (Quidel Quick View only)
- Cholesterol (Cholestech LDX, Chemtrak Accu-Check and Advance Care only)
- HDL cholesterol (Cholestech LDX only)
- Triglycerides (Cholestech LDX only)
- Screening test for Helicobacter pylori (Pyloritek Test only)
*Point-of-Care tests known to be performed at Yale-New Haven Hospital
|
Table 2
| Table 2. PROVIDER PERFORMED MICROSCOPY
|
|---|
| Moderately complex tests that are treated as waived tests if
the CLIA certificate lists only these and waived tests. See text)
- All direct wet mount preparations for the presences or absence of
bacteria, fungi, parasites, and human cellular elements.
- All potassium hydroxide (KOH) preparations
- Pinworm examinations
- Fern tests
- Post-coital direct, qualitative examinations of vaginal or cervical
mucous
- Urine sediment examinations
- Fecal leukocyte examinations
- Qualitative semen analysis (limited to the presence or absence of
sperm and the detection of motility).
|
Table 3
| Table 3. JCAHO REQUIREMENTS FOR WAIVED TESTING
|
|---|
- The organization has assessed the use of waived testing and
determined the extent to which it will be used
- The individuals responsible for performing waived testing and for
direction and supervision of the testing activities have been identified
- The staff performing waived testing have adequate, specific training
and orientation to perform tests and demonstrate satisfactory levels of
competence.
- For each waived test performed, there are specific policies and
procedures available to the person performing the test.
- Quality control checks are conducted for each procedure on a regular
basis as defined by the organization.
- Quality control records are properly documented and maintained.
|
Table 4
| Table 4. MAJOR REQUIREMENTS FOR MODERATELY COMPLEX TESTING
|
|---|
- Satisfactory participation in an accredited proficiency testing
program with at least three sets of unknowns per year.
- A documented Quality Control program with supervisory review of QC
results at least weekly
- Complete written procedures in compliance with National Committee for
Clinical Laboratory Standards (NCCLS) GP2-A2 document and available to
all testing personnel.
- An adequate system for reporting results, including establishing
critical values and documenting who performed each test.
- Proper labeling of reagents as to strength, storage requirements,
date prepared or reveived, date placed in service and expiration date.
- Documented criteria for determining the frequency of recalibration
- Use of two or more controls and evaluation of their results with each
analytical run
- A system for preventive maintenance and the regular checking and
documentation of operating characteristics for all instruments and equipment.
- Documented specific training in the test procedure and periodic
certification of competence for all testing personnel.
|
|