Time Delay Calibration
Purpose: The second laser on the FACSCalibur generates independent
data from the first laser and the machine must associate the two
datasets accurately in
order to provide correct profiles for each cell. This is done by calibrating
the time it takes for a cell to pass from the first laser to the second laser.
If the machine knows exactly how long this takes, it will be able to associate
the data from the first laser with the second laser, thus the name "time
delay calibration". There is no doubt that the machine can drift and it
may be out of calibration. Experience has shown that this will generate artifactual
data. In addition, if the machine cannot calibrate, this is an indication of
a serious problem, which should be corrected before you acquire data. Therefore,
it is recommended you perform this every time you intend to use the FL4 channel.
Description: The machine uses beads that fluoresce in channel 4 (e.g.Calibrite
APC Beads). When they pass by the first laser, they generate forward scatter
signals and when they pass by the second laser they generate fluorescence
signals. The machine adjusts its circuitry until these two signals are
correlated.
Materials:
APC Calibrite beads
Sheath fluid or PBS
FACS tubes
Procedure
- Shake the tube containing beads well- this is essential
as beads settle rapidly.
- Add 0.5ml of sheath/PBS to a FACS tube and then add one drop
of beads and mix.
- On the FACS machine, open the file "Time Delay Calibration" from
the desktop.
- Make sure the cytometer is connected to Cellquest (Acquire-Connect
to Cytometer).
- Open the following windows: Detectors/Amps, Threshold, Compensation
and Counters (Hold the Apple key down and then press 1, 2, 3, and 4).
- Make sure Four Color checkbox is selected at the bottom of the
Detectors/Amps window. The P7 parameter should change to FL4 in
the Detector column. P7 should be set to "Log" mode not "Lin". You will need
approx. 700-800 on the voltage setting.
- In the Threshold window, make sure Primary Parameter is FSC-H
(using "radio button"). Set the threshold for 200 using the slider pop up.
- Start the APC beads running, first on low pressure. In the Acquisition
panel, Setup should be checked: click Acquire. You may not see
anything at first. In the Detectors/Amps window, set the FSC gain up to around 6-7 or
whatever it takes to see events on the histogram. Note on the Time
Delay Calibration document window that there is a Histogram Statistic box. Adjust FSC gain until
the entire peak is above 400.
- Now increase the pressure to medium and then high. Note the shape of the histogram. It should not change much and should not
spread its width on high power. If it does, check for pressure leaks at the buffer supply area or at the tube attachment to the cytometer. If you are only getting a narrow spread on low or medium, you
will have to complete the Calibration on this pressure and will
have to run your experiment only at the lower pressure. The event
rate (in the Counters window) must be at least 400 under the pressure you are using. If not,
mix the beads well again and add another drop or remix the beads in the tube.
- Scroll down the Time Delay Calibration window to see the second
histogram clearly. It is of the FL4 channel. Now, adjust the gain
so that the mean peak in the M1 gate is at the "target value" of
the beads +/- 5. This can be seen on the label that is attached to the box. It is likely
that a value of 800 +/- will be sufficient.
- From the Cytometer Menu, choose Time Delay Calibration. After
a few seconds, the computer will beep and there will be no dialog
box. It only gives you feedback when there is a problem (i.e. signal out of range).
- If you didn't get an error, congratulations. You can now go on to set up the rest
of the machine using either beads, cells or both.