We have been
having severe problems with clogs on LSRIIs. This has
exacerbated problems with limited time slots as users cannot complete their
sample runs. In addition, it has required extra work to repeatedly unclog the
machines. The situation on the weekend has been particularly
unreliable.
To prevent clogs we will be enforcing the following rule as
of Feb 28th: All samples to be run on the LSRII must be filtered at the machine
just before running them. This can be accomplished with filter tubes or with
pieces of Nitex (see Analyzer rules for info). Violoators will be given a Level II violation. Repeat
offenders can lose privileges to use the facility and their PI will have to pay
fines. Please adhere to this new rule, which will be strictly enforced. We may
in the future extend it to the FACScaliburs, which are
also experiencing frequent clogs.
Please report any clogs or erratic
machine behavior as soon as you see it so we can take corrective action. Thanks
for your cooperation.