Troubleshooting This troubleshooting guide has also been placed on the desktop of all the FACS Facility computers. In the event of a problem with the machine, please seek Don Foster at TAC S617 or S631 for help. If he is not available please report your problem to him later via the form provided on top of the cytometers. You may also call Becton Dickinson for help. Their BD phone number is located on the side of FACS Calibur S along with the serial numbers of all the machines in the facility. They can walk you through most any problem. NOTE: The CellQuest Status window from Cytometer menu can be used to help diagnose STATUS problems. System Stays In NOT READY Check the following:
System Stays In STANDBY (System Not Pressurized) When the system is not pressurized, installation of a sample tube with the fluidics control button set to RUN will not change the status to READY it will remain in STANDBY. Pressure may be leaking from the reservoir caps or the vent valve, or the sample tube has not been pressurized properly. The sample either does not flow to the flow cell or flows poorly, and the FACSCalibur produces poor data. Check the following:
Bubbles are registered as events, producing spurious data. Bubbles can also cause alteration in the sample flow path, resulting in less than optimal data. The fluidic system should be re-primed. If the sheath reservoir was run dry, refill sheath and RUN system for 5 to 10 minutes with a test tube of distilled water before using for sample acquisition. This will remove bubbles and air from the sheath lines. No Events on the Computer Screen Check the following:
Sheath Fluid Dripping from the Sample Injection Tube Check the following:
|
||