FACS Aria* |
|
|
|
| |
|
|
|
| |
Violet Laser - 407nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
530/30 |
502LP |
Alexa Fluor 430, CFP |
|
|
| |
B |
450/40 |
|
Cascade Blue, Pacific Blue, DAPI,
Hoechst, Alexa Fluor 405, BFP, Marina Blue |
|
|
| |
|
|
|
| |
Blue Laser - 488nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
780/60 |
735LP |
PE-Cy7, PE-Alexa Fluor 750 |
|
|
| |
B |
695/40 |
655LP |
PI (with PE), PerCP, PerCP-Cy5.5,
PE-Cy5, EMA |
|
|
| |
C |
610/20 |
595LP |
PE-Texas Red, PI (with PE and PE-Cy5.5) |
|
|
| |
D |
576/26 |
556LP |
PE, PI (with PerCP) |
|
|
| |
E |
530/30 |
502LP |
FITC, GFP, CFSE, YFP, Alexa Fluor 488, SYBR-14 |
|
|
| |
F |
488/10 |
|
SSC |
|
|
| |
|
|
|
| |
Red Laser - 633nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
780/60 |
735LP |
APC-Cy7, APC-Alexa 750 |
|
|
| |
B |
710/25 |
685LP |
Alexa Fluor 680, Alexa Fluor 700 |
|
|
| |
C |
660/20 |
|
APC, Alexa Fluor 647 |
|
|
| |
|
|
|
| |
|
|
|
| |
*Please
consult FACS Facility Staff or http://www.bdbiosciences.com/spectra/ to |
|
|
| |
determine which
fluorochromes cannot be used together due to high spectral overlaps. |
|
|
|
|
| |
|
|
|
|
|
|
| |
LSRII* |
|
|
|
| |
|
|
|
| |
UV Laser - 355nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
530/30 |
450LP |
Indo-1 (Blue), Hoechst Red, Fura-2,
Am-Cyan |
|
|
| |
B |
440/40 |
|
Indo-1 (Violet), DAPI, Alexa Fluor
350, Hoechst |
|
|
| |
|
|
|
| |
Violet Laser - 405nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
525/50 |
505LP |
Alexa Fluor 430, CFP, Am Cyan |
|
|
| |
B |
440/40 |
|
Cascade Blue, Pacific Blue, DAPI,
Hoechst, Alexa Fluor 405, BFP, Marina Blue |
|
|
| |
|
|
|
| |
Blue Laser - 488nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
780/60 |
735LP |
PE-Cy7, Alexa 750 PE |
|
|
| |
B |
670/14 |
635LP |
PI (with PE), PerCP, PerCP-Cy5.5,
PE-Cy5, EMA |
|
|
| |
C |
610/20 |
600LP |
PE-Texas Red, PI (with PE and PE-Cy5.5) |
|
|
| |
D |
576/26 |
550LP |
PE, PI (with PerCP), Pynonin |
|
|
| |
E |
530/30 |
505LP |
FITC, GFP, CFSE, YFP, Alexa Fluor 488 |
|
|
| |
F |
488/10 |
|
SSC |
|
|
| |
|
|
|
| |
Red Laser - 633nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
780/60 |
735LP |
APC-Cy7, APC-Alexa Fluor 750 |
|
|
| |
B |
710/25 |
685LP |
Alexa Fluor 680, Alexa Fluor 700 |
|
|
| |
C |
660/20 |
|
APC, Alexa Fluor 647 |
|
|
| |
|
|
|
| |
|
|
|
| |
*Please
consult FACS Facility Staff or http://www.bdbiosciences.com/spectra/ to |
|
|
| |
determine which
fluorochromes cannot be used together due to high spectral overlaps. |
|
|
|
| |
|
|
|
|
|
|
| |
LSRII Green* |
|
|
|
| |
|
|
|
| |
Violet Laser - 407nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
525/50 |
505LP |
Alexa Fluor 430, CFP, Am Cyan |
|
|
| |
B |
450/50 |
|
Cascad Blue, Pacific Blue, DAPI,
Hoechst, Alexa Fluor 405, BFP, CFP |
|
|
| |
|
|
|
| |
Blue Laser - 488nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
670/14 |
635LP |
PerCP |
|
|
| |
B |
525/50 |
505LP |
FITC, GFP, PKH67, CFSE, Alexa Fluor 488 |
|
|
| |
C |
488/10 |
|
SSC |
|
|
| |
|
|
|
| |
Green Laser - 532nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
780/60 |
735LP |
PE-Cy7, |
|
|
| |
B |
710/50 |
685Lp |
PI,
PE-Cy5.5, |
|
|
| |
C |
660/20 |
630LP |
PE-Cy5 |
|
|
| |
D |
610/20 |
600LP |
Alexa Fluor 594, PE-Texas Red, Alexa
Fluor 568, PI, Rhodamine-B, PI (with PE and
PE-Cy5.5) |
|
|
| |
E |
575/26 |
|
PE, DsRed, RFP, Pyronin-Y,
PKH26, |
|
|
| |
|
|
|
| |
Red Laser - 633nm |
|
|
|
| |
|
|
|
| |
PMT |
Bandpass |
Long Pass |
Fluorochrome |
|
|
| |
A |
780/60 |
755LP |
APC-Cy7, APC-Alexa Fluor 750 |
|
|
| |
B |
710/50 |
690LP |
Alexa Fluor 680, Alexa Fluor 700 |
|
|
| |
C |
660/20 |
|
APC, Alexa Fluor 647 |
|
|
| |
|
|
|
| |
|
|
|
| |
*Please
consult FACS Facility Staff or http://www.bdbiosciences.com/spectra/ to |
|
|
| |
determine which
fluorochromes cannot be used together due to high spectral overlaps. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|